📞  +353 1 6643768
 
📧  info@medaval.ie
Rossmax CF175f (V1)
Device Model:
CF175f
Manufacturer:
Rossmax International Ltd., 12F - 189 Kang Chien Road, Taipei 114, TAIWAN ROC.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Device Accessibility:
Voiced instructions and results
Availability:
Available Currently
Device Manual:
Description:
The Rossmax CF175f (V1) is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a certified 4-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The accuracy of the technology used in the Rossmax CF175f (V1), to measure blood pressure, is certified by Medaval Ltd. It has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval ★★★★ Certified Accuracy Recent clinical validation; recent protocol with certified adherence
BP BIHS (UK and IRL) Self-measurement Published evidence
BP ESH (Europe) Self-measurement Published evidence
BP Stride BP Self-measurement for adults (Preferred) Published evidence
Device Family:
Validation Publications:

Zhang L, Kang YY, Zeng WF, Li Y, Wang JG. Validation of the Rossmax CF175 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2015 Apr;20(2):104-7. doi: 10.1097/MBP.0000000000000089. PMID: 25350783.

ESH-IP:2010 - Pass General population

Medaval Certification Assessment

Certification ImageMedaval Ltd. Accreditation assessment of the blood pressure measurement technology used in the Rossmax CF175 upper arm monitor, as validated according to the European Society of Hypertension International Protocol revision 2010. Medical Device Assessment. 2016 Aug 5;2016(1621) 5 p. Epub: 2019 Jan 31. Available from: www.medaval.ie.

With proven validity of validation results according to an accepted protocol, the clinical accuracy of the blood pressure measurement technology, used in the Rossmax CF-175F, is certified by Medaval Ltd.